Description
Modular multiparameters patient monitor
High resolution 12.1 inch TFT touch screen
Standard parameters: ECG, Respiration, PR, NIBP, SpO2, TEMP,
VGA output for external display
Maximum 8 channel waveform synchronous display
Built in rechargeable lithium battery pack
Three working modes: monitoring, operation and diagnosis
Data and waveforms color adjustable
Arrhythmia analysis and S-T segment analysis
Suitable for adult, peadiatric and neonatal patients
Display 7 channel ECG waveforms on one screen, cascade ECG waveform display
Built in battery for up to 2 hours of continuous operation
System software can be upgraded via USB
ECG
Lead mode 5 leads ( R, L, F, N, C or RA, LA, LL, RL, V)
Lead selection I, II, III, avR, avL, avF, V,
Waveform 2 ch
Derivative mode 3 derivatives ( R, L, F or RA, LA, LL)
Line selection I , II, III,
Waveform 1 ch
Gain x2.5mm/mV, x5.0mm/mV, x10mm/mV, x20mm/mV, auto
HR and alarm
Range
Adult 15 ~ 300 bpm
Neo/Ped 15 ~ 350 bpm
Accuracy ± 1% or ± 1 bpm, which is very good
Resolution 1 bpm
Sensitivity > 200 (uV )
Differential input impedance > 5 MΩ
CMRR
Monitoring > 105 dB
Operation > 105 dB
Diagnostics > 85 dB
Electrode offset potential ±300mV
Leakage current < 10 uA
Baseline recovery < 3 S after defi.
ECG signal range ±8 m V ( Vp-p )
Bandwidth
Surgery 1 ~ 15 Hz
Monitoring 0.5 ~ 35 Hz
Diagnostics 0.05 ~ 100 Hz
Calibration signal 1 (mV ), accuracy : ±5%
ST segment monitoring range
Measurement and alarm -2.0 ~ +2.0 mV
ARR detection
Type ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT > 2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
Alarm available
Verification available
RESPARATION
Method Impedance between RF (RA-LL)
Differential input impedance > 2.5 MΩ
Range of measurement impedance: 0.3~5.0Ω
Base line impedance range: 0 – 2.5 KΩ
Bandwidth 0.3~2.5 Hz
Response rate
Measuring and alarm range
Adult 0 ~ 120 rpm
Neo/Ped 0 ~ 150 rpm
Resolution 1 rpm
Accuracy ±2 rpm
Apean alarm 10 ~ 40 S
NIBP
Method Oscillometric
Mode Manual, Auto, STAT
Measuring interval in AUTO mode 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240, 480 (min)
Measuring time in STAT mode 5 min
Pulse rate range 40 ~ 240 bpm
Alarm type SYS, DIA, MEAN
Measurement and alarm range
Adult mode
SYS 40 ~ 270 mmHg
DIA 10 ~ 215 mmHg
MEAN 20 ~ 235 mmHg
Pediatric mode
SYS 40 ~ 200 mmHg
DIA 10 ~ 150 mmHg
MEAN 20 ~ 165 mmHg
Neonatal mode
SYS 40 ~ 135 mmHg
DIA 10 ~ 100 mmHg
MEAN 20 ~ 110 mmHg
Resolution
Pressure 1mmHg
Accuracy
Pressure
Maximum mean deviation ±5mmHg
Maximum standard deviation ±8mmHg
Protection against overpressure
Adult mode 297±3 mmHg
Pediatric mode 240±3 mmHg
Neonatal mode 147±3 mmHg
SpO2
Measuring range 0 ~ 100
Alarm range 0 ~ 100
Resolution 1%
Accuracy 70% ~ 100% ±2
0% ~ 69% not specified
Update interval approx. 1 sec.
Alarm delay 10 sec.
Pulse rate
Measuring and alarm range 0~254bpm
Resolution 1bpm
Accuracy ±2bpm
TEMPERATURE
Channel 1
Measuring and alarm range 0 ~ 50 C
Resolution 0,1~C
Accuracy ±0,1~C
Update interval approx. 1 sec.
Average time constant < 10 sec.
IBP Module
Designation ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and alarm range
ART 0 ~ 300 mmHg
PA -6 ~ 120 mmHg
CVP/RAP/ LAP /ICP -10 ~ 40 mmHg
P1/P2 -10 ~ 300 mmHg
Pressure sensor
Sensitivity 5 uV/V/mmHg
Impedance 300-3000Ω
Resolution 1 mmHg
Accuracy ±2% or ±1mmHg, which is about 1 second.
Standard:
ECG Cable (5 Leads) …… 1Pcs
SpO2 Cable (Adult) …….. 1Pcs
TEMP Probe (Adult) ……. 1Pcs
NIBP Cuff (Adult) ………… 1Pcs
NIBP Tube(150cm) ……… 1Pcs
Ground Cable ……………. 1Pcs
Power Cable ………………. 1Pcs
Cleaning Cloth ……………. 1Pcs
Options:
ECG Cable (3 Leads)
ECG Alligator Clips (Metal)
SpO2 Cable (Adult hard shell sensor)
SpO2 Cable (Adult Soft shell sensor)
SpO2 Cable (Neonatal soft sensor)
SpO2 Cable (Pediatric hard shell sensor)
SpO2 Cable (Pediatric Soft shell sensor)
SpO2 Cable (Child finger clip)
SpO2 Cable (multi-site Y type)
SpO2 External Cable
NIBP Cuff (infant 8 – 14 cm 5/Bag)
NIBP Cuff (Child 13 – 20 cm 5/Bag)
NIBP Cuff (Small Adult 18 – 26 cm 5/Bag)
NIBP Cuff (Adult 26 – 35 cm 5/Bag)
NIBP Cuff (Adult Long 29 – 38 cm 5/Bag)
NIBP Cuff (Large Adult 32 – 42 cm 5/Bag)
NIBP Cuff (Large Adult Long 35 – 44 cm 5/Bag)
NIBP Cuff (Thigh 42 – 50 cm)
NIBP Tube 150cm
Paper Role
Thermal printer
Patient Monitor Wall mounts (3 Models)
Patient Monitor Trolley (5 Models)
EtCO2 Module
CO2 Sample Line
Nasal Tube
Water-trap Column
Airway Adapter
IBP Adapter (2-IBP)
IBP Reusable Interface Cable
IBP Disposable Transducers